Herunterladen Inhalt Inhalt Diese Seite drucken

Packing; Passing On The Device - OHAUS Frontier FC5714 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für Frontier FC5714:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

EN-22
 Place the device on a stable, horizontal and non-resonant lab bench
1. Remove the front and back transport protection material.
2. Remove the plastic sleeve.
3. Observe a minimum distance of 30 cm to adjoining devices at the sides and
from the rear side to the wall.
4. Install the device in a well-ventilated location which is protected from direct
sunlight to prevent it from overheating.
 Connect the device
1. After installation, wait for four hours before switching the centrifuge on in order to avoid damage to the
compressor.
2. Check that the mains voltage and frequency match the requirements on the device name plate(see rear side of
the device) and then connect the device to the power supply.
 Remove the transport protection material from the rotor chamber
1. Switch on the device at the mains power switch.
2. Open the centrifuge lid using the open button.
3. Remove the transport protection material.
4. Place the rotor vertically onto the motor shaft.
5. Turn the rotor nut using the rotor key clockwise until the rotor nut is tightened.
 The device is now ready to use
Retain the packaging and all transport protection material for shipping the device at a later date.

8.3.3 Packing

Pack the centrifuge in reverse order.

8.3.4 Passing on the Device

When passing the equipment on to third parties, please make sure to also include this instruction manual.
Frontier
TM
Centrifuge FC5714/FC5718/FC5718R/FC5816/FC5816R/FC5916/FC5916R

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis