Safety instructions:
Special condition
Safety instructions:
Installation
Temperature tables
Connection data
8
• Avoid electrostatic charging of the plastic surfaces (especially at the version with outside diameter 29 mm).
Do not rub. Do not use in media or environments which may generate electrostatic charges on the plastic
surfaces.
• The mechanical strain level of the device corresponds to class "low" according to EN 60079-0, clause 26.4.2.
A
å 1
A
Zone 2
1
*
Only for versions with outside diameters 22 mm and 42 mm.
• Comply with the installation and safety instructions in the Operating Instructions.
• Install the device according to the manufacturer's instructions and any other valid standards and regulations.
• Do not operate the device outside the specified electrical, thermal and mechanical parameters.
• Connect cable screen to earth ground of the installation.
• The circuit of the device is isolated from ground potential and has a dielectric strength of at least 500 V
with respect to it.
When shortening the length of the cable ensure that this dielectric strength is maintained for connection
wires and earth grounded screen.
• Avoid impact and friction sparks (anchor equipment if necessary/secure against swinging).
• Do not disconnect electrical connections under voltage in an explosive atmosphere.
Ambient temperature range
–10 °C Ta  +70 °C
Max. ambient temperature
+70 °C
+60 °C
Maximum tensile load at the sensor
 100 N
Electrical data
U  30 V DC
(+)
(–)
1
*
Ingress protection
IP68
Temperature class
T5
T6
Cable material
PUR, PE, FEP
Waterpilot
FMX_03
rms