Once the analyzer warms up, the screen in (Figure 3) will appear offering the option to
change settings or start an analysis. If you wish to adjust factory 'Settings', select the left soft
key and refer to section
Maintenance & Troubleshooting. To begin an analysis, select
3
the right 'Start' soft key. Next select the type of refrigerant you wish to test (Figure 4).
NOTE:
If you are going to analyze a R-12 vehicle or cylinder you must select R-134a
mode.
Figure 3
Figure 4
2.3
Calibration
Each time the Legend ID™ begins a new test cycle it must complete an air calibration. The
calibration takes 30 seconds and pulls fresh air into the unit via an internal pump. This fresh
air purges any excess refrigerant from the unit and ensures accurate test results. Calibration
REQUIRES a sample hose be connected to the device and disconnected from the vehicle or
refrigerant source.
Once the sample hose is connected to the analyzer, press 'Start' to begin an air calibration,
as shown in (Figure 5). This will begin the calibration process and display the screen shown
in (Figure 6).
Figure 5
Figure 6
Manual Part Number:
Manual File:
Page 2-16
5-06-7000-80-0
012476 Rev. A