Dispensing
5.3

Changing the Syringe

The unit needs not be switched off to change the syringe.
Caution!
Press the hose clamp 11 closed so that pressure cannot be applies to the syringe
adapter 12 as a result of an unintentional start.
● Replace the Luer-Lock tip cap 17 on the filled syringe 15 with a dispensing needle 16.
● Replace the syringe and release the hose clamp 11.
To avoid air bubbles during dispensing, the tip of the syringe must be purged of air.
Caution!
Hold the syringe 15 over a container since product will flow out!
● Press button 6 CONT to switch to
continuous mode (see Section 5.2.2.2).
● Press and hold button 4 START (or the
Footswitch 20) until the product flows
free of bubbles from the dispensing
needle 16.
● Press button 6 CONT to switch off
continuous mode.
● Continue dispensing with the stored values or readjust the dispensed quantity
according to Section 5.2.2.
Caution!
For a change in the type of product, readjust the dispensed quantity according to
Section 5.2.2.
bar
START
CONT
psi
START
bar
CONT
psi
37