Guidance and Manufacturer's Declaration - Electromagnetic Immunity

LimbLogic is intended for use in the electromagnetic environment specified below. The
customer or the user of the LimbLogic should assure it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Note 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2:
These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cord-
DE-30
less) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and
Gebrauchsanweisung
IEC 60601
Compliance
test level
level
3 Vms
3 V
150 kHz to
80 MHz
3 V/m
3 V/m
80 MHz to
2.5 GHz
Electromagnetic
environment - guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the LimbLogic, including ca-
bles, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended
separation distance
d = 1.2√P
d = 1.2√P 80 MHz to 800 MHz
d = 2.3√P 800 MHz to 2.5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manu-
facturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,a should
be less than the compliance level in
each frequency range.b
Interference may occur in the vicinity
of equipment marked with the follo-
wing symbol:

Werbung