Herunterladen Inhalt Inhalt Diese Seite drucken

Operation With Other Supply And Evaluation Units - thyracont VCC200MA4 Betriebsanleitung

Vakuum transmitter
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

3.3.2 Operation with other Supply and Evaluation Units

The transducer can be operated with other customer related display units or volta-
ge supplies.
The electrical connection is to be made by means of suitable cables considering
EMI demands and according to the pin description shown below:
Socket
Pin1:
2
1
Pin2:
5
Pin3:
Pin4:
3
4
Pin5:
Maximum admissible load resistor in :
(Supply Voltage[V] - 9 V) / 0.02 A
Incorrect connection or inadmissible supply voltage can damage
the transducer!
M12, A-coded, 5pole, male
Do not connect
Signal Output 4-20 mA
Voltage Supply 9...30 VDC
Do not connect
Shield
VCC200-4-20 mA-201126
23

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis