Herunterladen Inhalt Inhalt Diese Seite drucken

Gima SUPERVEGA BATTERY Bedienungsanleitung Seite 23

Aspiratore
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 40
23
Guidance and manufacturer's declaration – electromagnetic Emissions
The SUPERVEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below.
The customers or the user of the SUPERVEGA BATTERY SUCTION UNIT should make sure that it's used in such an
environment.
Emissions test
Irradiated / Conducted
emissions CISPR11
Irradiated / Conducted
emissions CISPR11
Harmonic emissions
EN 61000-3-2
Voltage fluctuations / flicker
emissions EN 61000-3-3
Guidance and manufacturer's declaration – Immunity Emissions
The SUPERVEGA BATTERY SUCTION UNIT is intended for use in the electromagnetic environment specified below.
The customers or the user of the SUPERVEGA BATTERY SUCTION UNIT should make sure that it's used in such an
environment.
Immunity Test
Electrostatic
discharge (ESD)
EN 61000-4-2
Electrical fast
transient / burst
EN 61000-4-4
Surge
EN 61000-4-5
Loss of voltage, brief
voltage interruptions
and variations
EN 61000-4-11
Magnetic field with
network frequency
(50/60 HZ)
EN 61000-4-8
Note U
is the value of the power supply voltage
T
Compliance
Group 1
Class [B]
Class [A]
Complies
Level indicated by the
EN 60601-1-2
+/-8kV on contact
+/-15kV in air
± 2kV power supply
lines
± 1kV for input /
output lines
± 1kV differential mode
+/-2 kV ordinary mode
<5% U
(>95% dip U
)
T
T
for 0,5 cycle
40% U
(60% dip U
)
T
T
for 5 cycle
70% U
(30% dip U
)
T
T
for 25 cycle
<5% U
(>95% dip U
)
T
T
for 5 sec
30A/m
Electromagnetic environment - guidance
The SUPERVEGA BATTERY SUCTION UNIT only used RF energy
only for its internal functioning. Therefore, its RF emissions are very
low and are not cause interference in proximity of any Electronic ap-
pliances.
The SUPERVEGA BATTERY SUCTION UNIT can be used in all en-
vironments, including domestic and those connected directly to the
public mains distribution that supplies power to environments used
for domestic scopes.
Compliance Level
Electromagnetic environment - guidance
The device doesn't
Floors should be wood, concrete or ceramic
change its state
tile. If floors are covered with synthetic ma-
terial, the relative humidity should be at least
30%.
The device doesn't
Mains power quality should be that of a typical
change its state
commercial environment or hospital.
The device doesn't
Mains power quality should be that of a typical
change its state
commercial environment or hospital.
- -
Mains power quality should be that of a typi-
cal commercial environment or hospital If the
user of the SUPERVEGA BATTERY SUC-
TION UNIT request that the appliance oper-
ates continuously, the use of a continuity unit
is recommended.
The device doesn't
The power frequency magnetic field should
change its state
be measured in the intended installation lo-
cation to make sure that it's sufficiently low.
ENGLISH

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

28193

Inhaltsverzeichnis