1° case – If the overfill security system doesn't work the aspiration will be stopped by the bacteriological filter
who avoid the liquid penetration inside the device.
2° case – If both the security system doesn't work, there is the possibility that liquid comes inside the device,
in this case return the device to GIMA technical service.
GIMA S.p.A. will provide upon request electric diagrams, components list, descriptions, setting in-
structions and any other information that can help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
PLEASE CONTACT GIMA TECHNICAL SERVICE. GIMA DOES NOT GIVE GUARANTEE IF INSTRU-
MENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO BE TAMPERED.
INSTRUCTION FOR USE
Assembly of the device SUPERVEGA BATTERY SUCTION ASPIRATOR: Take the 5 arm base and set up
the 5 wheels that come with the above device. The wheels provided with braking device must be placed one
next to the order. Take the support bar that comes with the device SUPERVEGA BATTERY SUCTION ASPI-
RATOR and place it in the hole on the 5-arm base. From under the base, lock the two parts by means of the
supplied screw. Eventually, place the device on the trolley.
• The device must be checked before each use in order to detect malfunctions and / or damage caused by
transport and / or storage.
• The working position must be such as to allow one to reach the control panel and to have a good view of
the empty indicator, the jar and the antibacterial filter.
• If the device is to be transported from one place to another, to prevent the liquid collection jar from falling
and consequently the liquid from spilling, removing the jar from the device is recommended.
WARNING: For proper use, place the aspirator on a flat, stable surface, in order to have the full volume
of use of the jar and better efficiency of the overflow device. The vacuum jar, during use, must be used in
vertical mode, to prevent the action of the backflow valve. If this protection is triggered, turn the device off and
disconnect the pipe connected to the vacuum jar (indicated with the word VACUUM) on its cover.
Operation with AC/DC power pack:
• Connect the short silicone tube with the antibacterial filter to the suction port. The other tube, connected to
the filter on one end, must be connected to the spout on the vase lid with the float assembled inside (over-
flow device). The overflow device is triggered (the float closes off the internal lid fitting) when the maximum
volume (90% of the effective vase volume) is reached, thus preventing the liquid from penetrating the inside
of the machine. The device must be used on a flat, horizontal surface.
• Connect the long silicone tube to the free spout on the lid; the free end of the tube must be connected to
the conical fitting for the probe coupling, to which the suction probe must then be connected.
• Connect the universal power pack to the device using the dedicated connector and insert the power cord
plug into the socket. To start the treatment, press the switch into position I to turn the device on
• Set the desired vacuum level (Bar / kPa) through the vacuum regulator. Turn the knob in a clockwise direc-
tion to increase the vacuum level: these values can be read on the "vacuum gauge".
• To suspend and/or end the treatment, press the switch again and pull the plug out of the socket
• To mitigate the formation of foam inside the vase, unscrew and remove the lid from the vase, and fill the
latter with 1/3 water (to facilitate cleaning operations and speed up depressurisation during operation), then
screw the lid back onto the vase.
• Remove the accessories and proceed with cleaning operations.
• At the end of each use place the device back in the box, protected against dust
WARNING: The power cord plug is the element of separation from the electrical mains, even if the device is
equipped with an on/off button. Once the device is in use, the power plug must remain accessible to allow
another method of disconnection from the electrical mains.