Service information

Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to maintain
a device in accordance with the instructions may invalidate the device's warranty. Furthermore, failure to
maintain a device may compromise the clinical condition or safety of users and/or their caregivers. Do not
perform service and maintenance while the user is in the product.
1st Day
Read the User Guide carefully before use and
store it in a safe place for future reference
Store the enclosed tools on/at the product
Before every use
All nuts and bolts on this product must be
checked and tightened to avoid unnecessary
failures
Daily
Use a dry cloth to clean the product
Visual inspection. Repair or replace any
damaged or worn-out parts
Ensure that all fasteners are in place and fitted
correctly
Check all fittings and buckles for signs of worn
out parts
Be aware of any MAX mark indications before
adjustment
Wipe off any residue and dirt from the product
using a cloth with warm water and a mild
detergent/soap and dry before use
Weekly
Check that all castors are moving freely and
that all wheel locks/brakes are easily activated
Disinfect the product. please check the
disinfection section in this User Guide for more
information
Every 6 months
Check the condition of the upholstery and
replace it if necessary
Annually
Inspect the frame for cracks or signs for worn-
out parts and carry out annual service. Never
use a product which appears to be substandard
or faulty
etac.com
Washing
The cushion is removable and the cover can be
washed at max. 40°C using a mild detergent in a
washing machine. The upholstery is incontinence
proof and can be hand washed in warm water.
Warewashing machine
This product can be washed at 60°C using a mild
detergent in 15-20 minutes in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product. Clean the
accessories seperately.
Disinfection
The product without accessories can be disinfected
with a 70% disinfectant IPA solution. It is
recommended to wipe off any residue and dirt from
the product, using a cloth with warm water and a
mild detergent/soap without chlorine and let it dry
before disinfection.
Service interval
A detailed inspection of the product must be made
every 12 months (every 6 months for heavy users)
and every time the product is reissued for use. The
inspection should be carried out by a technically
competent person who has been trained in the use
of the product.
Service Check List:
Check and adjust the following functions as per their
intended use. Repair or replace any damaged or
worn-out parts. Please check the:
Fittings and buckles for signs of worn out parts
Castors are moving freely and all wheel locks/
brakes are easily activated
Nuts and bolts on the product are tightened
Product for cracks or signs for worn-out parts
All labels on the product are intact
23