8. Status- and Error-Codes
8.1 Logger Display
CAL
Calibration active
Strt
The logger is waiting for the recording start, according to the settings in the
datalogger setup menu.
StOP
Stop is reached if the logger is in the Start/Stop mode and the memory is full. Log-
ging has been stopped. For a new logging period, the logger has to be repro-
grammed. This is the delivery status of the PT100 loggers TP2 and TP4-L.
Ld.
The calibration parameters of the rH/T sensor are automatically uploaded when the
sensor is connected to the logger.
Prn
Data are printed
b.F.
Battery voltage is too low
CAL.E. Error during calibration
C.F.
Datalogger is faulty
Err
Error during printing
HHHH Logger keypad is defective
L.C.
Datalogger is faulty
Ld.F.
Error during upload of rH/T sensor parameters
n.c.
No sensor is connected
S.C.
Sensor has a short circuit
O.F.
Over flow of measurement value
U.F.
Under flow of measurement value
8.2 Logger Status in elproLOG ANALYZE
RAM IMG-BMP destroyed
This error message is visible in the logger status information, line: Module time. The reason for such an
error message might be, a battery replacement where the battery change time has not been set. (see also
chapter 2.7 Maintenance).
9. Hints
TH1 & TH2
Please do not calibrate with 2 connected rH/T sensors
For sensor exchange, the corresponding humidity channel (Channel 2 or 4) has to show
n.c. on the display at least twice before the new sensor is connected to the logger!
Please take care about the sensor position before connecting it. There is no force on the
sensor needed to plug it on. See also the printing on the casing of the logger (TH1 only)
TN2 & TN3-P Every time a data exchange with elproLOG software happens the function of the display
will be tested.
Digital Inputs In some of the documents the digital inputs are labelled with DigitIn. The same digital
inputs are labelled in the elproLOG software with D only.
Software
ECOLOG data loggers require an evaluation software of version 3.20 or higher.
CD-ROM
For more product information please see elproLOG ANALYZE CD-ROM.
EEx
All ECOLOG data loggers are
available with Ex Zone 1 approval.
For applications in hazardous areas the ELPRO documentation D-EZ-7003Ba
has to be taken into consideration!