Herunterladen Inhalt Inhalt Diese Seite drucken

MTR+ Rehalito Bedienungsanleitung Seite 64

Neuromuskulärer stimulator
Vorschau ausblenden Andere Handbücher für Rehalito:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

portable and mobile RF communications equipment and the Rehalito unit.
The Rehalito unit is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the Rehalito unit can help pre-
vent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the Rehalito unit as recommended
below, according to the maximum output power of the communications equipment.
Rated
maximum
output
150 kHz to 80 MHz
power of
transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Electromagnetic Compatibility
Recommended separation distances between
Separation distance according to frequency of transmitter
0.12
0.38
1.2
3.8
12
Rehalito Bedienungsanleitung / Operators Manual
m
80 MHz to 800 MHz
0.12
0.38
1.2
3.8
12
64
800 MHz to 2,5 GHz
0.23
0.73
2.3
7.3
23

Werbung

Kapitel

Inhaltsverzeichnis
loading

Verwandte Produkte für MTR+ Rehalito

Inhaltsverzeichnis