Herunterladen Inhalt Inhalt Diese Seite drucken

Advice On Application Of Electrodes - MTR+ Rehalito Bedienungsanleitung

Neuromuskulärer stimulator
Vorschau ausblenden Andere Handbücher für Rehalito:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Rehalito Bedienungsanleitung / Operators Manual

Advice on Application of Electrodes

Make sure, that the Rehalito is switched off (no display visible on the LCD-Display).
Connect the single poled ends of the lead wires to the electrodes plugs with the electrodes.
The output sockets of the electrode cables will be plugged into the channels A or B of the
Rehalito.
Furthermore make sure, that all plugs fit completely into the electrode sockets as well as the
device sockets.
Thereafter place the electrodes onto the skin. An adhesive skin gel has been applied to the
electrodes, which can be re-used after removal from the skin. In the case of the skin gel loosing
its adhesion and you can't achieve more than 6 mA - please replace the electrodes.
The same effect, however, also occurs when using a damaged cable. Therefore, if normal
treatment is not possible by using new electrodes, the cable maybe faulty.
Electrodes which are damaged due to careless handling may not be used.
Electrodes with damaged cable insulation (e.g. visible copper wire) may
not be used.
Electrodes with damaged, missing or stained aluminium coating or carbon
layer, also with insufficient adhesive gel may not be used.
Electrodes may not be altered or tampered with (e.g. by scissors or other tools).
Do not pull the electrodes by force on the electrode cable. To remove the electrodes from the
skin, please lift the electrode at the rim towards the middle.
Do not use electrodes with ointments or creams (especially which enhance the blood flow or
act as local anaesthesia).
Apply the electrodes on clean and healthy skin only.
In the case of bodily hair growth, the hair should be shortened with scissors.
Neither stretch the skin nor the electrodes when applying the electrodes.
Effective current densities of more than 2 mA/cm
require an increased attention of the user
2
with regard to the intensity adjustment.
For standard treatment we recommend the use of our self-adhesive electrodes with an area of
40x80 mm.
If you should use a probe, please clean it carefully before and after use and dry it completely.