Display and Operating Elements
„On/Off": By pressing the button once, the device will either be switched on or off.
„+/- buttons": These buttons adjust the intensity for channels A and B. The device has two
independently controllable output channels A and B. When pressing the buttons the intensity
will either be increased or decreased in steps of 1 mA. 60 seconds after the last change of
intensity, a further increase may not be carried out due to an electronic barrier. Pressing the „+"
button three times releases the barrier and the intensity can be altered again.
A reduction of the intensity is possible at any time.
Whilst configuring program PC05 one switches between the programmable parameters by
pressing the left „+/-" button (channel A), and through pressing the right „+/-" button (channel
B) the desired value/data for respective parameter can be set. This way, the application time, the
stimulation-mode, the frequency and the pulse width can be selected.
„P": This button selects preset programs 01–04 or custom program PC05.
Press the „P"-button once at any time during a running program to reduce the mA on both
channels to zero and pause the program. Pressing „P" a second time ends the program
Holding „P" for three seconds while statistics are displayed will exit the statistics screen without
deleting them. Rehalito will return to the home screen of the selected program and the lock will
be cancelled.
The „P" button stores all adjustments made to the custom program mode and returns to the
home screen.
„S": This button is used to setup custom program 05. Press the „S"-button for at least
3 seconds in program PC05 to enter the custom program setting. Pressing „S" a second time for
10 seconds will enter the real time clock (RTC) setup.
Press the „S" button once at any time during a running program, to reduce the mA on both
channels to zero and pause the program. Pressing „S" a second time will resume the program
and the current will increase over a period of one second back to the previously set intensity
level.
Holding the „S" button for 5 seconds while the statistics are displayed, will display „dEL".
Holding „S" a second time for 5 seconds will delete the statistics, set the custom program to
default values and show the RTC setup screen. Turn off the unit if you do not want to delete the
statistics.
The unit is intended to be supported by hand during normal use. The belt clip is for temporary