Guidance and manufacturer's declaration – electromagnetic emissions
The Rehalito unit is intended for use in the electromagnetic environment specified below.
The customer or the user of the Rehalito should ensure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Rehalito unit is intended for use in the electromagnetic environment specified below. The
customer or the user of the Rehalito should ensure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Electromagnetic Compatibility
Compliance
Group 1
Class B
Not Applicable
Not Applicable
IEC 60601
Compliance
test level
+/-6 kV contact
+/-6 kV contact
+/-8 kV air
+/-8 kV air
3 A/m
3 A/m
Rehalito Bedienungsanleitung / Operators Manual
Electromagnetic environment – guidance
The Rehalito uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any interfe-
rence in nearby electronic equipment.
The Rehalito is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Electromagnetic environment –
level
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.
62