Herunterladen Inhalt Inhalt Diese Seite drucken

Description Of The Functions Of The Training - MEDICA THERA-vital Gebrauchsanweisung

Vorschau ausblenden Andere Handbücher für THERA-vital:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Description of the functions of the training

Training time
The training time can be preset from 1 to
60 minutes in every training program. Stand-
ard is 30 minutes.
Automatic change of direction of rotation
THERA-vital changes every 4 minutes in
passive training the direction of rotation.
While training actively, there will be no chang-
ing of the direction of rotation.
Sensitivity of spasm control
This feature enables you to set the borderline
for the sensory of THERA-vital to notice a
wrong reaction of the patient that will cause
the engine to stop, in order to minimize any
possible overstraining.
For patients tending to show any symp-
toms of spasm choose the adjustment as
follows:
fine
medium for patients reacting less sensible.
rough
Downloaded from
www.Manualslib.com
Spasm program
The spasm-feature of THERA-vital sets in
which direction the device starts again after a
spasm.
Choose the adjustment as follows:
Forward: when having the tendency for
stretch spasm;
Backward: when having the tendency for
bowing spasm.
Automatic Number of Revolutions-Adjust-
ment
At the active training the adjustment of the
number of revolutions automatically adjusts
to a higher capacity level of the patient, if the
automatic adjustment of the number of revo-
for patients usually reacting sensi-
lutions is set "on".
bly. With this adjustment usually a
low resistance is sufficient to stop
If the patient works with a number of revolu-
spasm.
tions of 30 rpm instead of the preset number
of revolutions of i.e.15 rpm, THERA-vital will
adjust after approximately 10 sec the preset
for patients with a low risk for
number of revolutions on 25 rev./min. If the
spasm. The device will stop with a
patient afterwards gets passive again,
high resistance.
THERA-vital keeps working with 25 rev./min.
This way the patient is working towards his
own individual number of revolutions and
when showing passiveness will get trained
with a convenient number of revolutions.
manuals search engine
The sensibility is also closely related
Motor power
to the radius of the handle (see
The preset motor power has a stronger effect
pages 31, 32) and the engine power
on the patients legs if a small pedal radius is
(see below).
chosen as if a big radius is chosen.
Active muscle-support
This function helps patients that are unable to
get a smooth rotation with just one leg or arm.
The motor supports the active side by periodi-
cally doing the work on the in-active side
needed. The feature works like a flywheel at
mechanical training devices, only that
THERA-vital does not have to get driven. Any
observers will not notice that the patient only
works one-sided. Is the patient able to work with
both sides the active muscle-support is not to
notice.
39
Please set the motor power as follows:
high when pedal radius is big
(100 % motor power),
medium when pedal radius is medium
(approx. 80 % motor power),
low when pedal radius is small
(approx. 60 % motor power).
Small pedal radius and adjustment of
motor power on "big" can effect very
big powers on the patient.

Werbung