9619 NB / 9627 NB HD Monitor
Elektromagnetische Verträglichkeit (EMV)
Guidance and manufacturer's declaration – electromagnetic immunity
9619 NB/9627 NB is intended for use in the electromagnetic environment specifi ed below.
The customer or user of the 9619 NB/9627 NB should ensure that it is used in such an environment.
Immunity test
EN/IEC 60601 test level
Electrostatic Discharge (ESD)
± 6 kV contact
IEC 61000-4-2:2008
± 8 kV air
Electrical fast
± 2 kV for power supply lines
transient/burst
± 1 kV for input/output lines
IEC 61000-4-4:2004
Surge
± 1 kV differential mode
IEC 61000-4-5:2005
± 2 kV common mode
Voltage dips, short
interruptions and voltage
<5 % U
variations on power supply
(>95 % dip in U
input lines IEC 61000-4-11
for 0.5 cycle
<40 % U
(60 % dip in U
for 5 cycles
<70 % U
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 sec
Power frequency (50/60 Hz)
magnetic field
3 A/m
IEC 61000-4-8:2004
* Note: U
is the a.c. mains voltage prior to application of the test level.
T
9619 NB / 9627 NB HD Monitor
Electromagnetic Compatibility (EMC)
Compliance level
Complies
± 6 kV contact
± 8 kV air
Complies
± 2 kV for power supply lines
± 1 kV for input/output lines
Complies
± 1 kV differential mode
± 2 kV common mode
Complies
*
<5 % U
*
T
T
)
(>95 % dip in U
T
for 0.5 cycle
Complies
<40 % U
T
T
)
(>60 % dip in U
T
for 5 cycles
Complies
<70 % U
T
T
)
(>30 % dip in U
T
for 25 cycles
Complies
<5 % U
T
T
)
(>95 % dip in U
T
for 5 sec
Complies
3 A/m
9619 NB / 9627 NB Monitor HD
Compatibilidad electromagnética (CEM)
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30 %.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
)
If the user of the equipment or system requires
T
continued operation during power mains interruptions,
it is recommended that the equipment or system be
powered from an uninterruptible power supply.
)
T
)
T
)
T
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.