Elektromagnetische Verträglichkeit (EMV)
Guidance and manufacturer's declaration – electromagnetic immunity
9619NB/9626NB(-2)/9627NB(-2) is intended for use in the electromagnetic environment speci ed below. The customer or user of the
9619NB/9626NB(-2)/9627NB(-2) should ensure that it is used in such an environment.
Immunity test
EN/IEC 60601 test level
Electrostatic Discharge (ESD)
± 6 kV contact
IEC 61000-4-2: 2008
± 8 kV air
Electrical fast
± 2 kV for power supply lines
transient/burst
± 1 kV for input/output lines
IEC 61000-4-4: 2004
Surge
± 1 kV differential mode
IEC 61000-4-5 : 2005
± 2 kV common mode
Voltage dips, short
interruptions and voltage
<5% U
variations on power supply
(>95% dip in U
input lines
for 0.5 cycle
IEC 61000-4-11
<40% U
(60% dip in U
for 5 cycles
<70% U
(30% dip in U
for 25 cycles
<5% U
(>95% dip in U
for 5 sec
Power frequency
(50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8: 2004
* Note: U
is the a.c. mains voltage prior to application of the test level.
T
9619 NB / 9626 NB(-2) /
9627 NB(-2) – HD Monitor
Electromagnetic Compatibility (EMC)
Compliance level
Complies
± 6 kV contact
± 8 kV air
Complies
± 2 kV for power supply lines
± 1 kV for input/output lines
Complies
± 1 kV differential mode
± 2 kV common mode
Complies
*
<5% U
*
T
T
)
(>95% dip in U
T
for 0.5 cycle
Complies
<40% U
T
T
)
(>60% dip in U
T
for 5 cycles
Complies
<70% U
T
T
)
(>30% dip in U
T
for 25 cycles
Complies
<5% U
T
T
)
(>95% dip in U
T
for 5 sec
Complies
3 A/m
9619 NB / 9626 NB(-2) /
9627 NB(-2) – HD
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
)
If the user of the equipment or system requires
T
continued operation during power mains interruptions,
it is recommended that the equipment or system be
powered from an uninterruptible power supply.
)
T
)
T
)
T
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
(
)