Herunterladen Inhalt Inhalt Diese Seite drucken

Guidance And Manufacturer's Declaration - Electromagnetic Emissions - Storz 9619 NB Gebrauchsanweisung

19/27 hd monitor
Vorschau ausblenden Andere Handbücher für 9619 NB:
Inhaltsverzeichnis

Werbung

37
9619 NB / 9627 NB HD Monitor
Elektromagnetische Verträglichkeit (EMV)
Guidance and manufacturer's declaration – electromagnetic emissions
9619 NB/9627 NB is intended for use in the electromagnetic environment specifi ed below.
The customer or user of the 9619 NB/9627 NB should ensure that it is used in such an environment.
Therefore, its RF emissions are very low and are not likely to cause interference in nearby electronic equipment.
Emissions test
Radiated emission CISPR 22
EN 55022:2006 +A1:2007
30 MHz ... 1 GHz - Class B
1 GHz ... 6 GHz - Class B
Conducted at mains terminal
EN 55022:2006 +A1:2007 –
0.15 MHz ... 30 MHz Class B
Harmonic emissions
IEC 61000-3-2:2005
+A1:2008 +A2:2009
Voltage fluctuations/
flicker emission
IEC 61000-3-3:2008
9619 NB / 9627 NB HD Monitor
Electromagnetic Compatibility (EMC)
9619 NB/9627 NB uses RF energy only for its internal function.
Compliance
9619 NB/9627 NB is suitable for use in all establishments (e.g. hospitals and
Class B
medical practices) including domestic establishments and those directly
30 MHz ... 6 GHz
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Class B
0.15 MHz ... 30 MHz
100 Hz ... 2 kHz
P
≤ 1.0; P
≤ 1.0;
st
eT
d
≤ 3.3 %; d
≤ 4 %;
c
max
d
≤ 3.3 % (t=500 ms)
(t)
9619 NB / 9627 NB Monitor HD
Compatibilidad electromagnética (CEM)
Electromagnetic environment – guidance

Werbung