Herunterladen Inhalt Inhalt Diese Seite drucken

Pari COMPACT 2 Gebrauchsanweisung Seite 118

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 49
CAUTION
Quality of treatment impaired by electromagnetic
interference
Electrical devices can cause electromagnetic interference.
Interference can impair the function of the devices and thus
also the effectiveness of the treatment.
• Do not place the PARI device immediately beside or on
top of other devices.
• Keep at a minimum distance of 30 cm from portable wire-
less communication devices (including accessories there-
for, such as antenna cables or external antennas).
• If the PARI device has to be placed immediately beside or
on top of other devices for operation, all devices must be
monitored to ensure that they are working properly.
CAUTION
Danger of injury from falling compressor
A compressor placed in a poor position poses a risk of injury.
• Do not place the compressor above head height.
• Make sure that it cannot be pulled down by the power
cord or the connection tube.
• Do not place the compressor on a soft surface such as a
sofa, a bed or a tablecloth.
NOTE
Danger of device fault caused by dust
If the compressor is operated in a very dusty atmosphere,
more dust may collect inside the housing. This may cause a
fault in the device.
• Do not operate the compressor on the floor, under the bed
or in workshops.
• Only operate the compressor in a low-dust environment.
– 18 –

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis