The Soft Start mode may be deselected/selected only
when the pump is stopped.
If necessary press CYCLE NUMBER to select YES or
NO, then enter the selection by pressing the PUMP
CURRENT push-button. The display shows:
F R O N T
/
R E M O T E
S E L E C T
I O N : X X X X X X
where: XXXXXX = means the word FRONT or
REMOTE, or RS 232 depending on the last
selection.
The controller is factory-set for FRONT panel
operation.
After choosing the desired selection by pressing the
CYCLE push-button, press the PUMP CURRENT
push-button to enter the value.
The display shows:
R S 2 3 2
B A U D
R A T E
where: XXXX = means 600, 1200, 2400, 4800, 9600
baud
rate
for
the
communication. The controller is factory-set for 9600
baud rate operation.
If necessary, select the desired value by pressing the
CYCLE NUMBER, then enter the value by pressing the
PUMP CURRENT push-button.
NOTE
/
2 3 2
X X X X
host
computer
or
printer
The display shows:
R S 2 3 2
S E L E C T
where: XXXX == means HOST or PRINT.
Select HOST or PRINT by pressing the CYCLE
push-button.
59
TECHNICAL INFORMATION
H O S T
/
P R I
I O N :
X X X X X
87-900-869-01 (B)