N.
108
109
110
111
117
120
121
122
123
124
125
TV 401/301/21 NSF User Manual / 87-900-995-01 (C)
RS 232/RS 485 Communication Description
Read/Write
Datatype
R/W
N
W
L
R/W
L
R/W
L
R/W
N
R/W
N
R/W
N
R/W
L
Reserved to Agilent service
R/W
L
Technical Information
Description
Baud rate
Pump life/ cycle time/ cycle
number reset
Interlock type (default = 1)
Analog output type: output
voltage signal proportional to
frequency or power
Low Speed frequency (Hz)
Rotational frequency setting
(Hz)
Maximum rotational frequency
in Hz (active only in Stop
condition)
Set vent valve on/off
(on = closed)
Set the vent valve operation
17
Admitted Values
600 = 0
1200 = 1
2400 = 2
4800 = 3
9600 = 4
(default = 4)
To reset write '1'
Impulse = 0
Continuous = 1
0 = frequency
1 = power
(default = 0)
750 to
"Maximum
rotationl
frequency"
(win 121)
(default = 750)
750 to
"Maximum
rotationl
frequency"
(win 121)
(default = 1010)
750 to 1010
(default = 1010)
On = 1
Off = 0
(default = 1)
Automatic = 0
(see note 1.)
On command = 1
(see note 2.)