Herunterladen Inhalt Inhalt Diese Seite drucken

Medtronic MINILINK REAL-Time Bedienungsanleitung Seite 7

Vorschau ausblenden Andere Handbücher für MINILINK REAL-Time:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 33
low, the charger will not work. Repeat the battery installation
steps using a new battery.
Charging the MiniLink
Before using the MiniLink for the first time, you must fully charge the
MiniLink battery which may take up to 8 hours. It is recommended
to recharge the MiniLink after each sensor use. If you choose to
recharge the MiniLink after a 3-day sensor use, the charging time
will be less than 20 minutes. A fully charged MiniLink battery will
work more than 14 days without recharging. After 14 days of use, the
MiniLink will fully recharge in less than 2 hours.
1. If a green light on the MiniLink is lit or flashing, do not connect
it to the charger. The MiniLink will not charge with its green
light on. Wait for the green light to turn off (approximately 30
seconds), then connect the MiniLink to the charger.
2. Connect the MiniLink to the charger
by lining it up, flat side down, with the
charger. Push the two components
together fully.
3. Within 10 seconds after the MiniLink
is connected, a green light on the
charger will flash for 1-2 seconds as
the charger powers on. For the rest of
the charging time, the charger's green
light will continue to flash in a pattern
of four flashes with a pause between
the four flashes.
4. When charging is complete, the green
light on the charger will stay on,
without flashing, for 15-20 seconds and then turn off.
5
After the green charger light turns off,
disconnect the MiniLink from the charger.
The green light on the MiniLink will flash for
about 5 seconds and then turn off.
green light
- 5 -

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis