INFECTION CONTROL PROCEDURES
Warnings
• Instruments are not sterile upon receipt and must be sterilized
prior to use in accordance with the following instructions.
• If an ultrasonic bath is used, the insert should be bathed
separately and singularly to prevent the diamond-coated tip
from damaging other instrumentation.
• The use of cleaning and disinfectant solutions is not recommended.
Limitations on Processing
• Repeated cleaning and sterilization cycles have minimum
effect on these instruments. End of life is normally determined
by wear and damage due to use.
• Do not use chemical disinfectants prior to sterilization or rapid
deterioration of the material may occur.
• Cold liquid disinfection/sterilization, chemical vapor sterilization,
and dry heat sterilization methods have not been tested or
validated for efficacy and are not recommended for use.
Instructions
Point of use:
Remove excess soil with disposable cloth or paper.
Containment and Transportation
Protect instruments from damage. Instruments should be
reprocessed as soon as is reasonably practical following use.
Instruments left standing wet may stain or corrode.
Cleaning: Automated
• May be cleaned using an ultrasonic water bath for 15 minutes.
Refer to Warnings before using an ultrasonic bath.