DIRECTIONS FOR USE
• Sterilize before using. Refer to the Infection Control Procedures.
• Copious water flow, at least 35ml per minute, should always be
maintained when using Diamondcoat inserts.
• Ensure water reaches diamond-coated tip prior to use.
• It is recommended that periodic evacuation and brief air
drying of the site be carried out to assess the progress of
a given procedure.
• Clinical experience should guide the operator to determine the best
power adjustment for a particular procedure. Most procedures will
be effectively accomplished with a Medium power setting however,
the full range of system power from low to high can be used.
• Use a smooth, continuous motion and light contact with the
Diamondcoat inserts. Heavy pressure should never be used.
Excessive pressure may result in possible soft tissue or tooth
4
surface damage and excessive wear of the diamond coating.
INSERT MARKING
17081 30K DIAMONDCOAT
Manufacturer, Date (YYDDD=Double digit year and triple digit
day of year), Frequency, Description, Lot Number.
O-RING REPLACEMENT
Before placing into the handpiece, performing the following
steps will increase the o-ring life on the Diamondcoat inserts:
• Fill entire handpiece with water
• Lubricate o-ring (black rubber material) with water.
• Gently twist the insert down into the handpiece until the insert
is fully seated in place.