Injection Transformer User Manual
WARNING
Death or serve injury due to hazardous voltage levels.
Use proper safety probes and make sure that the injection transformer and safety
probes are connected to safety ground using a proper ground connection when
working on voltage levels >30 V
CAUTION
Minor or moderate injury in case of equipment failure.
Always connect the injection transformer's earth terminal to safety ground using a
proper ground connection when voltage levels >30 V
setup.
NOTICE
Do not apply DC signals to the output of the injection transformer.
Otherwise the injection transformer will be damaged!
10
-
B-WIT 100 contains a 1 A fuse (not replaceable by user)
-
B-LFT 100 contains a 100 mA fuse (not replaceable by user)
or >60 V
AC
DC
AC
or >60 V
are present in the
DC
OMICRON Lab