Herunterladen Inhalt Inhalt Diese Seite drucken

Gas Ballast; Bearing Purge; Purge Kit; Vpi Valve Installation Data - Agilent Technologies TriScroll 600 Serie Installations- Und Betriebshandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 13
Wiring
Using the data in Table 19, verify that the VPI Valve
chosen is compatible with the TriScroll supply voltage
and frequency. Then, locate the proper wiring diagram to
use (Figure 4 on page 37 or Figure 5 on page 38).
1. Remove the three screws (item 12 in Figure 3 on
page 36) that are holding the motor electrical cover
(item 13 in Figure 3).
.
Motor
TriScroll Vacuum Pump
Operating Voltage
1 Phase
100-120 V 50/60 Hz
1 Phase
200-230 V 50/60 Hz
3 Phase
200 V 50/60 Hz
3 Phase
230 V 50/60 Hz
3 Phase
380 V 50/60 Hz
3 Phase
415 V 50/60 Hz
3 Phase
460 V 50/60 Hz
*
Solenoid operating voltages are lower than the TriScroll vacuum pump operating voltages in order to utilize the TriScroll motor's inter-
nal thermal switch to actuate the VPI Valve.

Gas Ballast

The pump incorporates an automatic gas ballast to
prevent water and other condensates from accumulating
within the pump. The standard configuration is a sintered
filter installed in the 1/4" National Pipe Thread gas ballast
port (item 9 on Figure 3 on page 36). This allows enough
atmospheric air to enter the pump in order to purge
condensates while not affecting pump ultimate pressure
or pumping speed.
For applications where the ingress of atmospheric air is
undesirable, dry nitrogen at a flow rate of
bled into the gas ballast port. See "Purge Kit" below.
Blocking of the gas ballast port is not recommended.
TriScroll 600 Series Vacuum Pump
Table 19

VPI Valve Installation Data

VPI Valve Part #
VPI Valve Part #
NW25
VPI251205060
VPI401205060
VPI251205060
VPI401205060
VPI251205060
VPI401205060
VPI251335060
VPI401335060
VPI252205060
VPI402205060
VPI252205060
VPI402205060
VPI252665060
VPI402665060
5 lpm can be
2. Connect the VPI Valve solenoid wire leads to the
pump motor as shown in Figure 4 or Figure 5.
3. Replace the motor electrical cover and secure it using
the three screws removed in step 1. Verify that the
valve is properly grounded before applying
electrical power.
Solenoid Operating
NW40
Voltage Range
90-132 V
90-132 V*
90-132 V*
120-146 V*
180-264 V*
180-264 V*
239-293 V*

Bearing Purge

A 1/4" National Pipe Thread bearing purge port (item 6 on
Figure 3 on page 36) protects the main crankshaft
bearings in applications where large amounts of water are
being pumped. In the standard configuration, this port is
sealed.
To enable the bearing purge, dry nitrogen at a flow rate of
5 lpm can be bled into the bearing purge port. This gas
supply should be maintained at 2 psig or less and must be
kept below 5 psig. See "Purge Kit" below.
CAUTION
The bearing purge port has been
disabled on the TriScroll 600
model S4800309. Use of this port
could cause internal damage to
the pump.

Purge Kit

A purge kit (Agilent part number PTSPURGEKIT) is
available to properly purge either the bearing purge or the
gas ballast. This kit contains a flowmeter and all
necessary valving and tubing.
Use the following
Wiring Diagram
Figure 4 low voltage
Figure 4 high voltage
Figure 5 low voltage
Figure 5 low voltage
Figure 5 high voltage
Figure 5 high voltage
Figure 5 high voltage
41

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis