4.
Pay attention to all areas where fluids or soil may gather, such as
under/ around the handles or any tight crevices/ grooves.
5.
Dry the unit and/or accessories with a dry, white lint-free cloth.
6.
Conduct a final confirmation of the cleaning process by visually
inspecting the white cloth for remaining soil.
7.
If soil remains on the white cloth, repeat steps 3 through 6.
8.
After cleaning is complete, turn the unit on to perform Power On
Self-Test (POST). If any errors are received, contact AtriCure to
begin return process.