AtriCure CRYO1 Installationsanweisungen Seite 7

Zylinderdruckregulierer
Vorschau ausblenden Andere Handbücher für CRYO1:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 31
2.
Remove the clamp from the Cylinder Pressure Regulator by completely
removing the two screws with the provided 5/64" allen wrench.
3.
Push the Cylinder Pressure Regulator shaft all the way into the desired Probe
Supply Socket Connector. There should be no shaft showing when completely
inserted.
4.
Attach Cylinder Pressure Regulator to Supply Socket Connector with clamp.
Align the grooves and capture with clamp
Align the two clamp pieces by matching up the larger hole with the smaller hole.
Tighten the clamp down with the two screws using the allen wrench.
5.
The Cylinder Pressure Regulator Unit installation is complete.
NOTE: The color bands around the System ports are for identification and
reference only. Actual color may vary from Green that is shown in this
instruction.