2. CCS–200 Shut Down
NOTE: For additional help on shutting down the system, please refer to the
frigitronics CCS-200 Operation and Maintenance Manual.
1. Turn the Nitrous Oxide Cylinder Valves clockwise to the "OFF"
position.
Turn the system On/Off Valves "ON" and "OFF" repeatedly until all
pressure gauges read "0" psi and the Exhaust Pressure Gauge needles
touch the silver stop.
2. Turn system On/Off Knobs to the "OFF" position.
CAUTION: Do not remove or install probes unless the line and exhaust
pressure gauge read "0" psi.
3. If attached, remove probe(s).
4. Turn Main Meter switch to the "OFF" position.
5. The CCS–200 shut down is complete.

3. Replacement versus First Time Installation

NOTE: If this is a first time installation of the Cylinder Pressure Regulator,
skip Section 4 and proceed directly to Section 5.