Herunterladen Inhalt Inhalt Diese Seite drucken

User Maintenance; Fuse Replacement; Technical Service - Braun Aesculap GN200 Gebrauchsanweisung/Technische Beschreibung

Lektrafuse hf-generator
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 22
7.6

User maintenance

In case of error, follow the instructions in the display and note the error
code.
If the error continues after following the instructions in the display,
contact your national B. Braun/Aesculap agency, see Technical Service.
7.7

Fuse replacement

Risk of fatal injury from electric shock!
Unplug the device before changing the fuses!
DANGER
Specified fuses: 2 units Aesculap art. no. TA021404:
G fuse link, time lag (T) 8 A, breaking capacity H (1 500 A)
Use a small screwdriver to release the clip on the fuse holder 11 and
tilt up cover.
Use a small screwdriver to release the fuse holder 11 and pull it out.
Replace both fuses.
Reinsert the fuse holder 11.
Close the cover.
Fig. 12
Note
If the fuses burn out frequently, the device is faulty and should be
repaired, see Technical Service.
8.

Technical Service

Risk of injury and/or malfunction!
Do not modify the product.
WARNING
For service and repairs, please contact your national B. Braun/Aesculap
agency.
Modifications carried out on medical technical equipment may result in
loss of guarantee/warranty rights and forfeiture of applicable licenses.
Service addresses
Aesculap Technischer Service
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone:
+49 (7461) 95 -1601
Fax:
+49 (7461) 14 -939
E-Mail:
ats@aesculap.de
Or in the US:
Aesculap Inc.
Attn. Aesculap Technical Services
615 Lambert Pointe Drive
Hazelwood
MO, 63042 USA
Aesculap Repair Hotline
Phone:
+1 (800) 214 -3392
Fax:
+1 (314) 895 -4420
Other service addresses can be obtained from the address indicated above.
17

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis