Lektrafuse HF Generator GN200
Validated cleaning and disinfection procedure
Validated procedure
Wipe disinfection for electrical
devices without sterilization
5.6

Wipe disinfection for electrical devices without sterilization

Phase
Step
I
Wipe disinfection
RT:
Room temperature
Phase I
Remove any visible residues with a disposable disinfectant wipe.
Wipe all surfaces of the optically clean product with a fresh, disposable
disinfectant wipe.
Observe the specified application time (1 min minimum).
5.7

Inspection, maintenance and checks

Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
Set aside the product if it is damaged.
Perform a safety inspection annually, see Maintenance.
5.8

Storage and transport

Damage to the product caused by switch-on too
early after storage/transport at temperature below
+10 °C (50 °F)!
CAUTION
Allow the HF device to acclimate at room tem-
perature for about 1 hour.
Transport the product only in its original box.
For storage and transport conditions, see Ambient conditions.
12
®
Special features
T
t
Conc.
[°C/°F]
[min]
[%]
≥1
RT
-
Water quality
Chemical
-
Meliseptol HBV wipes 50 % Propan-1-ol
6.

Maintenance

No servicing is required for the Lektrafuse HF generator GN200.
Note
Maintenance must only be carried out by authorized personnel. If neces-
sary, the circuit diagrams and the service manual, which contains all the
necessary documents, are provided for this.
6.1

Safety inspection

A safety inspection must be carried out annually.
The person in charge of the inspection will document the inspection find-
ings and measurement values according to the printed inspection proto-
col.
Only have the product and its accessories inspected by persons who
possess the requisite training, knowledge and experience and who are
authorized to do so.
If there are significant deviations from the inspection protocol values,
or if the prescribed maximum values are exceeded: Send in the device.
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
Reference
Chapter Wipe disinfection for electrical devices
without sterilization