4. Filters
a. Ensure filters are securely seated.
b. Perform the ULPA filter test (Section 4.2) prior to the start of each procedure.
5. Tubing
a. Check throughout the procedure to be sure tubing is not occluded.
Caution:
from being aspirated/pulled into the tubing.
3.4
Clean-Up
After each procedure, the SurgiFresh® Turbo smoke evacuation system should be cleaned and
prepared for future use.
1. Turn off the power switch and unplug the unit from the wall receptacle.
2. Remove the unit from the immediate procedure area.
3. Wipe the footswitch and footswitch tubing with an appropriate hospital disinfectant. Coil the
footswitch tubing around the cord wrap at the back of the unit.
4. Wipe the power cord with an appropriate hospital disinfectant. Coil the power cord around
the cord wrap.
5. Remove the tubing from the pre-filter or suction canister.
6. If a suction canister was used, securely close lids/caps and properly dispose canister.
7. Remove the pre-filter.
8. Dispose of all operative waste materials according to your institution's procedures for
biohazardous materials.
9. Check the ULPA filter to be sure no fluids are on top of the filter. Replace if fluids are
present.
10. Install a new pre-filter.
11. Thoroughly wipe all external surfaces of the SurgiFresh® Turbo with an appropriate hospital
disinfectant. Follow the procedures approved by your institution.
Warning:
4.
MAINTENANCE
4.1 Power Cord
1. Prior to each use, check the power cord along its entire length and at both plugs to ensure
no damage has occurred.
2. Do not use a power cord with exposed wires, cracks, or frayed areas.
4.2 Filter Test
Note:
1. Ensure an ULPA filter is in the unit and seated properly.
OPERATING INSTRUCTIONS
Ensure sponge guard is attached to prevent surgical materials or tissue
Ensure the unit is completely dry before restoring power.
MAINTENANCE
Perform the ULPA filter test with no pre-filter or tubing in place.
8