OPERATING INSTRUCTIONS
2.
OPEN PROCEDURES -- EVACUATING SMOKE WITH INCIDENTAL FLUIDS
a. Ensure ULPA filter / pre-filter are installed and tested per section 2.1
b. Secure a disposable suction canister near the unit.
c. Attach a 1 1/3" diameter hose from the suction canister to the pre-filter.
Note: Hose attaches to the 'vacuum' port on a suction canister.
d. Connect the adapter and/or tubing appropriate for the procedure to the 'patient' port of the
suction canister.
Caution:
If 3/8" tubing is used, avoid dislodging from the suction canister by using a
sufficient length for the procedure.
3.
CLOSED PROCEDURES
Note: Closed procedure set up for dry smoke or smoke with incidental fluid is the same as
the corresponding open procedure set-up.
Warning: Some cannulas, valves, or other instruments combined with suction may
cause rapid reduction of the pneumoperitoneum in laparoscopic procedures.
6