Recommended separation distances between portable and mobile RF communications
The MCC-500MD is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the MCC-500MD can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the MCC-500MD as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Caution
When you dispose of the unit or accessories,
you must obey the laws in the relative area or
country and the regulations in the relative
hospital.
Warning
Using this unit for medical purposes
This equipment's connectors are not
isolated.
Do not connect any device other than one
which conforms to IEC60601-1.
When an information technology device or
AV device that uses an alternating current is
connected, current leakage may result in an
electric shock to the patient or operator.
If use of such a device is unavoidable, isolate
its power supply by connecting an isolation
transformer, or by connecting an isolator
between the connecting cables.
After implementing these measures, confirm
that the reduced risk now conforms to
IEC60601-1.
equipment and the MCC-500MD
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
m
80 MHz to 800 MHz
d = 1.2√P
0.12
0.38
1.2
3.8
12
Caution
Do not come into contact with the terminals of
the rear panel connectors and patients at the
same time.
Doing so may result in a generation of
voltage that can be harmful to patients if the
unit is malfunctioning. Always disconnect the
power cord before connecting and
disconnecting connectors.
For the customers in Europe
This product has been manufactured by or
on behalf of Sony Corporation, 1-7-1 Konan
Minato-ku Tokyo, 108-0075 Japan.
Inquiries related to product compliance
based on European Union legislation shall be
addressed to the authorized representative,
Sony Deutschland GmbH, Hedelfinger
Strasse 61, 70327 Stuttgart, Germany.
For any service or guarantee matters, please
refer to the addresses provided in the
separate service or guarantee documents.