Herunterladen Diese Seite drucken

Agilent Technologies TwisTorr 84 FS AG Bedienungshandbuch Seite 163

Vorschau ausblenden Andere Handbücher für TwisTorr 84 FS AG:

Werbung

PROGRAMMABLE ANALOG SIGNAL: this output signal is a voltage
(from 0 to 10 Vdc) proportional to a reference quantity (frequency or
power or Temperature or Pressure) set by the user. The default
setting is the frequency (see the following example diagram).
Figure 16
Type
Frequency
Power
Temperature
Pressure
TwisTorr 84 FS AG Navigator Controller User Manual / 87-901-038-01
10
0
0
Analog Output Voltage proportional to frequency (WIN 111=0)
Range
0 ÷ 1350Hz
0 ÷ 100W
0 ÷ 100 °C
1*10
[Torr] ÷ 1*10
-9
Technical Information
frequency [Hz]
Output Analog Voltage
0 ÷ 10Vdc
0 ÷ 10Vdc
0 ÷ 10Vdc
[Torr]
0 ÷ 10Vdc
-1
17
J1 – IN-OUT
1350
WIN 111
0
1
2
3
163/192

Werbung

loading

Diese Anleitung auch für:

X3509-64000X3509-64001