Herunterladen Inhalt Inhalt Diese Seite drucken

Motorised Saddle Height Adjustment; Connecting The Blood Pressure Cuff; Connecting The Device - Ergosana Sana Bike 350 F Gebrauchsanweisung

Fahrradergometer
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

3.2.3

Motorised saddle height adjustment

The device is available with an optional motorised saddle height adjustment. At the rear side of the
control console, there are two switches, showing an up and down arrow. By means of these switches,
the saddle height is infinitely variable. The maximum upper and lower saddle position is fixed. The
adjustment range covers heights between 1.20 and 2.10 meters. The maximum patient weight is 160
kg.
Caution:
When adjusting the saddle to a higher position, it is recommended to relieve the hoist motor by shift-
ing the body weight onto the pedals.
Saddle up
3.2.4

Connecting the blood pressure cuff

The interfaces for the air tube and the microphone are located at the bottom of the control panel. The
air tube is connected via a special plug coupling. To connect or disconnect the tube, pull back the
sleeve shell. The microphone socket is located next to the air tube interface. (Warning! Observe the
marking).
3.2.5

Connecting the device

Establish potential equalisation (see section 2.3) and plug the supplied power cable into an earthed
socket. As the device is preset to the local mains voltage, you can switch it on using the main switch
on the rear panel. The ergometer is now ready for operation.