Technical data

Note
SITRANS TF
For SITRANS TF, please refer to the built-in transmitter's specifications (SITRANS TH200, TH300, TH400)
Input
Interface
1)
● Standard connection
● Averaging
● Differentiation
Sensor current
Measuring cycle
Break monitoring
Short circuit monitoring
For connection diagrams see ConnectionHotspot text (Page 11)
1)
Resistance thermometers
Table 7-1
Measured variable: Temperature
Input point
According to IEC 60751
● Pt25
● Pt50
● Pt100 ... Pt200
● Pt500
● Pt1000
● Ni25 ... Ni1000
According to JIS C1604-81
● Pt25
● Pt50
● Pt100 ... Pt200
● Pt500
● Pt1000
● Cu10 ... Cu1000
26
SITRANS TH100
x
approx. 0,4 mA
< 0,7 s
SITRANS TH100/200/300,
TR200/TR300
(for TH100, only PT100), min. span
10 °C (18 °F)
Range °C (°F)
-200 ... +850 (-328 ... +1562)
-200 ... +850 (-328 ... +1562)
-200 ... +850 (-328 ... +1562)
-200 ... +850 (-328 ... +1562)
-200 ... +350 (-328 ... +662)
-60 ... +250 (-76 ... +482)
-200 ... +649 (-328 ... +1200)
-200 ... +649 (-328 ... +1200)
-200 ... +649 (-328 ... +1200)
-200 ... +649 (-328 ... +1200)
-200 ... +350 (-328 ... +662)
SITRANS TH200/TH300,
TR200/TR300
x
x
x
≤ 0,45 mA
≤ 250 ms for 1 sensor with
break monitoring
x
x
Range °C (°F)
-200 ... +850 (-328 ... +1562)
-200 ... +850 (-328 ... +1562)
-200 ... +850 (-328 ... +1562)
-200 ... +850 (-328 ... +1562)
-200 ... +850 (-328 ... +1562)
-60 ... +250 (-76 ... +482)
-200 ... +850 (-328 ... +1562)
-200 ... +850 (-328 ... +1562)
-200 ... +850 (-328 ... +1562)
-200 ... +850 (-328 ... +1562)
-200 ... +850 (-328 ... +1562)
-50 ... +200 (-58 ... +392)
SITRANS TH400
x
x
x
nominal 0,2 mA
≤ 400 ms
x
x
SITRANS TH400
SITRANS TH/TR/TF
A5E32363963-01, 06/2013