Operating Parameters
84
Parameter 11: Interval between two latches per timer. The
following values can be set directly by the timer of the IK 220:
Parameter
value
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
32
Parameter 12: To permit a time interval of over 3.2 milli-
seconds, parameter 12 can realize a counter that uses only
every nth timer pulse for latching. To permit a latching interval
of 9 milliseconds, for example, parameter 11 must be set to 46
(corresponds to 3 milliseconds) and parameter 12 set to 3.
1)
Only possible without compensation of encoder signals and
without calculation of compensation values.
2)
With compensation; without calculation of compensation
values.
Time
Parameter
interval
1)
100 µs
2)
150 µs
200 µs
250 µs
300 µs
350 µs
400 µs
450 µs
500 µs
550 µs
600 µs
650 µs
700 µs
750 µs
800 µs
900 µs
Time
value
interval
33
1000 µs
34
1100 µs
35
1200 µs
36
1300 µs
37
1400 µs
38
1500 µs
39
1600 µs
40
1800 µs
41
2000 µs
42
2200 µs
43
2400 µs
44
2600 µs
45
2800 µs
46
3000 µs
47
3200 µs

Quicklinks ausblenden: