Regulations and declaration of manufacturer- Electromagnetic compatibility, interference
emission
The Précise 6000 is intended for use in an environment as described below. The operator of the
Précise 6000 must make sure that the unit is operated in such an environment.
Immunity test
Discharge of static
electricity (ESD)
acc. to
IEC/EN 61000-4-2
fast transient
electric
interference /
bursts acc. to
IEC 61000-4-4
Surge voltage acc.
to
IEC/EN 61000-4-5
Voltage dips,
shortterm
interruptions and
fluctuations in
supply voltage acc.
to IEC 6100-4-11
Magnetic field at a
supply frequency
(50/60Hz) acc. to
IEC 61000-4-8
Note: Ut is the a.c. supply voltage before application of the test levels.
Rev. D 2018-11
IEC/EN 60601-
test level
+/- 6kV contact
discharge
+/- 8kV
Air discharge
+/- 2kV for mains
leads
+/- 1kV for input /
output leads
+/-1kV Push - pull
+/-2kV asymme-
trical voltage
>5% Ut
(>95% dip in Ut) for
½ period
40% Ut
(60% dip in Ut) for
5 period
70% Ut
(30% dip in Ut) for
25 period
>5% Ut
(>95% dip in Ut) for
5 s
3 A/m
Compatibility
level
+/- 6kV contact
discharge
+/- 8kV Air
discharge
+/- 2kV for mains
leads
+/- 1kV for input /
output leads
+/-1kV Push - pull
+/-2kV asymme-
trical voltage
>5% Ut
(>95% dip in Ut)
for ½ period
40% Ut
(60% dip in Ut) for
5 period
70% Ut
(30% dip in Ut) for
25 period
>5% Ut
(>95% dip in Ut)
for 5 s
3 A/m
Electromagnetic environment -
regulations
Floors should be made of wood or
concrete or should be covered with
ceramic the floor is covered with
synthetic materials, the relative
humidity must be at least 30%.
The quality of the supply voltage
should match the typical office or
hospital environment.
The quality of the supply voltage
should match the typical office or
hospital environment.
The quality of the supply voltage
should match the typical office or
hospital environment.
Magnetic fields at mains frequency
must comply with the typical vales
found in office or hospital
environments.
32