Only qualified service personnel may check the apparatus in case of damages to
the plug or mains cable, malfunctioning, general damages from falling down
onto the ground or into water.
In case the apparatus cannot be switched off, pull out the plug and call a service
technician.
Smoking, open fire and light are not allowed when using the device.
The user is not put at risk by external electromagnetical interferances.
Portable and mobile RF communication equipment can affect medical
electrical equipment. This may cause a malfunction of the device.
For this reason, radiation sources (devices emitting electromagnetic waves) in the
vicinity of medical electrical equipment have to maintain a given minimum distance. See
following table.
Radiation sources
Mobile phone
(cell phone)
DECT phone
(cordless phone)
Bluetooth devices
(laptops, cell phone)
Radio remote controls
WLAN devices
(e.g.: laptops, repeater,
Access Point, Print
Server)
The power cord must not exceed a maximum length of 4m.
The use of accessories and cables other than those listed in the instructions for use may
lead to increased interference emissions or reduced immunity to interference of the
device.
Rev. D 2018-11
Minimum distance from medical
electrical equipment, including all
connecting cables
3.3 m
1.2 m
0.7 m
1.2 m
2.3 m
23