3. Diagnosis
An extensive autodiagnosis system has been implemented into the oxygen concentrator
6000. By pressing the button next to the oxygen outlet on the right the diagnosis
Précise
is called up. The displays can be called up by pressing the button.
Servicebutton
press x times
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
Rev. D 2018-11
Designation
Language ENGLISH
Softwareversion
<<V1.56 –M->>
Serial number
O2 Sensor value
Flow Sensor value
Power switch inverts
Concentration monitoring:
< 80% O2
Concentration monitoring:
< 72% O2
Battery monitoring
O2 < 82% Ionization disconnection
O2 Concentration display
Flow to small (Monitoring)
Flow to high (Monitoring)
Equipment version
Service telephon number
Interval time
Date
Time
Working hours counter
min. of pressure values of all 3
columns
max. of pressure values of all 3
columns
Error memory bank of last 10
Alarms
Information
Select language
Software version & equipment
e.g. Précise 6000M
Individual serial number on the
equipment
Temperature and
concentration value
Flow value l/min
Factory adjusted
Factory adjusted
Factory adjusted
Factory adjusted
Factory adjusted
Factory adjusted
Factory adjusted
Factory adjusted
Factory adjusted
Servicetelefonnummer des
Fachhandelspartners
Factory adjusted
Adjusted
Adjusted
1 = Patient 2 = Total
Defect of main valve, columns,
sensor, compressor etc.
Defect of main valve, columns,
sensor, compressor etc.
Indicates the error with date
and time
26