With the serial communication you can use the following
windows:
Win 224
XX
Win 257
Win 267
The windows from 136 to 140 manage an additional set point
(set point 3) dedicated to turn ON or OFF the gauge (win 267=4)
TwisTorr 84 FS AG Rack Controller User Manual / 87-901-037-01 (A)
Programmable Analog Output Diagrams
Pressure reading with the format X.X E
Gauge status (read only):
0
when a gauge is not connected
1
when a gauge is connected
2
when the pressure read is under
range or when the gauge give an
internal error
3
when the pressure read is under
range or when the gauge give an
internal error
4
when the "identification" is present
but not known by the controller
Gauge power (read/write):
If you need to turn off the gauge at
certain condition you can use this
window
0
gauge off
1
gauge on (default)
2
gauge normally off. It turns on when
the set point 1 is activated. If the set
point 1 type (win 101) is selected as
"pressure" the gauge is always on.
3
gauge normally off. It turns on when
the set point 1 is activated. If the set
point 2 type (win 171) is selected as
"pressure" the gauge is always on.
4
gauge normally off. It turns on when
the set point 1 is activated. If the set
point 3 type (win 101) is selected as
"pressure" the gauge is always on.