Cleaning
Clean the weighing tray and the housing as
required using a domestic cleaning agent or a
commercially-available disinfectant. Follow
the manufacturer's instructions.
What do I do if ...
... no weight is displayed when there is a
weight on the scale?
Is the scale switched on?
Check the batteries.
0. 0 00
...
does not appear before
weighing?
Switch the scale off and back on again,
making sure there is no load on the scale.
...
appears in the display?
Battery power is dropping. Change the
batteries in the next few days.
bAtt
...
appears in the display?
The batteries are completely discharged.
Insert new batteries.
St0p
...
appears in the display?
The maximum load has been exceeded.
Er:011:
...
appears in the display?
The scale is overloaded or the load is
unevenly distributed. Take the load off the
scale or distribute weight more evenly and
restart scale.
Er:012:
...
appears in the display?
The scale was switched on with too great a
load on it. Remove load and restart.

Technical data

Dimensions
Depth:
Width:
Height:
Weight
Temperature range
+10 °C to +40 °C
Maximum load
Graduations
Up to 10 kg / 22 lbs:
Up to 20 kg / 44 lbs:
Accuracy
better than ±0.3% / ±10 g
Auto Hold function
Power supply
mains unit (option)
Power consumption
approx. 30 mA
Accessories
Plug-in mains unit
300 mm
Use only seca original plug-in mains units with
640 mm
9 V or controlled 12 V output voltage.
105 mm
Commercially-available mains units may
approx. 3.1 kg
supply a higher voltage than stated and thus
lead to damage to the scale.
20 kg / 44 lbs
Disposal
5 g / 0.5 ozs
Disposing of the device
10 g / 0.5 ozs
Do not dispose of the device in
domestic waste. The device must be
yes
disposed of properly as electronic
battery/
waste. Follow the national regulations which
apply in your case. For further information,
contact our service department at:
service@seca.com
Batteries
Do not dispose of used batteries in household
waste. Dispose of batteries via local collection
points. When buying new supplies, buy
batteries low in harmful substances which
contain no mercury (Hg), cadmium (Cd) or
lead (Pb).