12. Specifications
12. Specifications
Model CTD9100-1100
Indication
EN
Display
Temperature range
1)
Accuracy
2)
Stability
Display resolution
Temperature distribution
3)
Axial homogeneity
4)
Radial homogeneity
Temperature control
Heating time
Cooling time
5)
Stabilisation time
Temperature block
Dimensions (D x H)
Block depth
Insert
Immersion depth
Insert dimensions
Insulator dimensions
Insert material
Voltage supply
Power supply
Power consumption
Fuse
Power cord
Communication
Interface
Case
Dimensions (W x D x H)
Weight
1) Is defined as the measuring deviation between the measured value and the reference value.
2) Maximum temperature difference at a stable temperature over 30 minutes.
3) Maximum temperature difference at 40 mm [1.57 in] above the bottom.
4) Maximum temperature difference between the bores (all thermometers inserted to the same depth).
5) Time before reaching a stable value.
The measurement uncertainty is defined as the total measurement uncertainty (k = 2), which contains the following shares:
accuracy, measurement uncertainty of reference, stability and homogeneity.
Certificates
Certificate
Calibration
Recommended recalibration interval
Approvals and certificates, see website
For further specifications see WIKA data sheet CT 41.29 and the order documentation.
20
2-line LC display 20 ch x lines (3.2 x 5.5 in) with backlighting
200 ... 1,100 °C [392 ... 2,012 °F]
±3 K
±0,4 K at 1,000 °C [2,012 °F]
0.01 °C up to 999 °C, then 0.1 °C [0.01 up to 1,830 °F, then 0.1]
dependent on temperature, temperature probes and their quantity
dependent on temperature, temperature probes and their quantity
approx. 50 min (approx. 18 °C/min) from 20 °C to 900 °C [from 68 °F to 1,652 °F]
approx. 150 min from 1,100 °C to 200 °C [from 2,012 °F to 392 °F]
dependent on temperature and temperature probe
Ø 44 x 300 mm [Ø 1.73 x 11.81 in]
175 mm [6.89 in]
155 mm [6.10 in]
Ø 42.5 x 175 mm [Ø 1.67 x 6.89 in]
Ø 42.5 x 65 mm [Ø 1.67 x 2.56 in]
Ceramic
AC 230 V, 50/60 Hz or AC 110 V, 50/60 Hz
950 VA
6.3 A slow blow fuse
AC 230 V; for Europe
RS-232
170 x 390 x 330 mm [6.69 x 15.35 x 12.99 in]
12 kg [26.46 lbs]
Standard: 3.1 calibration certificate per DIN EN 10204
Option: DKD/DAkkS calibration certificate
1 year (dependent on conditions of use)
WIKA operating instructions, model CTD9100-1100