7. Operating of the calibrator
Function
Units °C/°F/K
DEF.PAR. ON/OFF
EN
KEY

7.2.4 Third level - Operation settings

Following functions can only be reached, if the „KEY" in the second level is entered over the buttons [▲] / [▼] is correct and
then the buttons [M] and [▲] have to be pressed simultaneously. Back over simultaneous pressing of the buttons [M] and [▲]
or waiting 20 s.
Function
Access key
Baud rate
Address
S/N
Board S/N
MAX. SET.
MIN. SET.
WAIT 0/1
REV. SOFTWARE
SENSOR TYPE
STAB: +/-0.3 °C
16
Meaning
Setting of the unit, which should be displayed
Default Parameter
The controller can handle default and customized settings of P.B./T.I./T.D.
OFF: customised settings
ON: default settings
The controller is adapted to the calibration ex-factory. If there are any change wishes,
please contact the manufacturer.
Do not change these settings!
Code for the functions in the third level (Default 2)
Meaning
Access code
(Default = 2)
Velocity of data transfer
(Default = 9600)
Communication address
(Values from 1 to 32 possible)
Serial number of the instrument
(not editable)
Serial number of the board
(not editable)
Maximal temperature set point
(not editable)
Minimal temperature set point
(not editable)
Setting "0": Calibrator starts to control the set point after switching it on
Setting "1": Calibrator starts to control after switching it on, when the tester has confirmed
Firmware Version
Type of internal reference sensor
Symbol ÷ appears on the display, if the temperature is stable < +/- 0, 3 °C over 6 minutes.
WIKA operating instructions, model CTD9100-1100