Herunterladen Inhalt Inhalt Diese Seite drucken

Operating Of The Calibrator; Setting A Temporary Set Temperature (Set-Point Mode); Programming (Main Menu) - WIKA CTD9100-1100 Betriebsanleitung

Vorschau ausblenden Andere Handbücher für CTD9100-1100:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6. Commissioning, operation / 7. Operating of the calibrator
After the test or calibration
WARNING!
Risk of burns!
High temperatures can lead to acute burns.
At the end of the calibration DO NOT remove
the probe, if the block is still indicating high
temperature.
Cool down the calibrator including the test
item in order to avoid a thermal shock.
Before switch off the calibrator make sure
that the temperature of the block is almost the
same as ambient temperature.
No
> 1 mm
Yes
Fig. 1

7. Operating of the calibrator

7.1 Setting a temporary set temperature (set-point
mode)
Set the required temperature value:
Pressing the button [▲] to increment the set point.
Pressing the button [▼] to decrement the set point.
The button [E] confirms the entry.
Wait for the stabilisation of the set point temperature before
starting any calibration (symbol ÷ on the first line of the
display).

7.2 Programming (Main menu)

All the settings can be carried out in this menu structure.
1. Press button [M].
This opens the main menu.
2. With the button [M] to select the desired entry in the main
menu (see overview).
WIKA operating instructions, model CTD9100-1100
No
Reference
probe
Fig. 2
Calibration zone
Fig. 3
3. With the button [E] confirm the selected menu item.
The calibrator has three function levels:
First level: general settings
Second level: specific functions for optimising the
controller
Third level: operation settings
EN
No
Yes
13

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis