Error
Cause
description
IR communication error
Err 4
Power pack or battery
voltage too low.
Err 5
Battery contacts are
corroded.
Air congestion
Err 6
Blood pressure cuff
incorrectly connected.
Leak in the blood
pressure cuff or air hose
Measurement cancelled
Abr. (Abort)
on request
Remedy
If this error message is displayed continuously (or repeatedly),
contact your healthcare provider or the manufacturer.
➔
Change the batteries (
Chapter 5.2).
Clean the battery contacts with a cotton cloth and a little alcohol.
Check the blood pressure cuff for air congestion or a kink in the air
hose. If there is a kink in the air hose, unkink it, otherwise send the
blood pressure monitor to your healthcare provider or the
manufacturer.
Connect the blood pressure cuff to the blood pressure monitor
➔
(
Chapter 6.2).
Replace the blood pressure cuff with the air hose.
Do not press this button when measuring, unless you want to abort
the measuring.
Error messages
EN
39