Important!
Do not wash the device PierenPlus basic with water. Do not clean it
with gas, diluter, or other solvents. Please take care that no liquid enters the device.
If, however, liquid did enter the device, it must undergo a technical check before
being reused.
Troubleshooting
Problem
Possible Cause
»
The device cannot
be switched ON.
»
»
The intensity can-
not be set above
»
4 mA.
»
The device
switches OFF
suddenly.
»
The stimulation is
not felt, or the sti-
»
mulation feeling is
not sufficient
»
»
The batteries are not or not
correctly inserted
The batteries are weak or empty
One or both electrodes are not or
not correctly placed on the skin
The cables are not correctly
connected to the device or to the
electrodes
One or both electrodes are not in
place or fell off the skin
The electrodes are not put cor-
rectly onto the skin
The electrodes are put too close
to each other on the skin
The set intensity is not high
enough
The batteries are too weak
Suggestion to Solution
»
Insert new batteries in the right direction
»
Insert new batteries or accumulators
»
Check for correct connection of the
cables and electrodes on the skin
»
Connect the parts correctly
»
Check all cables
»
Check the correct placing of the elec-
trodes
»
Check the electrodes for correct placing
»
Place the electrodes so that contact
between them is avoided minimum 2 cm
distance
»
Raise the intensity with the [▲] keys, until
the stimulation is well felt, but not painful
or causing discomfort
»
Insert new batteries or accumulators
English