Technical data on electromagnetic compatibility (EMC instruc-
tions) in table format are included with the compressor. It can
also be ordered from PARI GmbH or downloaded from the
internet at www.pari.de/en/products under "Technical Data" on
the respective product page.

6.4 Recommended safety distances

Recommended safety distances between portable and mobile
high-frequency (HF) telecommunication devices and the PARI
compressor:
The PARI compressor is intended for operation in an
electromagnetic environment in which radiated HF transients are
monitored. The user can help to prevent electromagnetic
interference by observing the minimum safety distances
between portable and mobile HF communication equipment
(transmitters) and the PARI compressor, as recommended
below in accordance with the maximum output power of the
communication device.
Safety distance depending on transmitting frequency:
Rated power
of the
transmitter [W]
0.01
0.1
1
10
100
- 51 -
150 kHz to 80
MHz
D = (3.5/3) ×
SQRT(P) [m]
0.2
0.4
1.2
3.8
12
80 MHz to 800
MHz
D = (3.5/3) ×
SQRT(P) [m]
0.2
0.4
1.2
3.8
12
800 MHz to
2.5 GHz
D = (7/3) ×
SQRT(P) [m]
0.3
0.8
2.3
7.3
23