Technical data
Measuring ranges
and resolutions
with oxygen
sensor, CellOx 325
Measuring ranges
and resolutions
with oxygen
sensor, DurOx 325
Accuracies
Correction
functions
30
Measuring ranges, resolutions, accuracies
Variable
Concentration [mg/l]
Saturation [%]
T [°C]
Variable
Concentration [mg/l]
Saturation [%]
T [°C]
Variable
Concentration [mg/l]
Saturation [%]
T [°C]
Temperature
compensation
Salinity correction
Air pressure
correction
Measuring range
0 ... 19.99
0 ... 90.0
0 ... 199.9
0 ... 600
0 ... 50.0
Measuring range
0 ... 90.0
0 ... 600
0 ... 50.0
Accuracy (± 1 Digit)
± 0.5 % of measured value
at ambient temperature of 5 °C ... 30 °C
± 0.5 % of measured value
when measuring in the range of ± 10 K around the
calibration temperature
± 0.1
Accuracy better than 2 % at 0 ... 40 °C
0 ... 70.0 SAL
Automatic through integrated
pressure sensor in the range 500 ... 1100 mbar
Oxi 1970i
Resolution
0.01
0.1
0.1
1
0.1
Resolution
1
1
0.1